PREPARE TEST CARD
CLOSE DEVICE
7. Read Results
Remove card from its pouch just before use and lay flat.
9. Ordering and Contact Information
Immediately peel off adhesive liner from the right edge of the test card. Close and securely seal the card. Gently snap off swab shaft, which is purposely scored. Avoid dislodging the swab from the well when doing so. Do not reopen.
COLLECT SPECIMEN—NASOPHARYNGEAL SWAB
EndFragment
READ NOW RESULTS WITH ALERE READER
STANDARD
URINE COLLECTION
CONTAINERS
ALERE BINAXNOW®
LEGIONELLA
URINARY
ANTIGEN CARD
DEMO
In Read Now mode, set timer and wait 15 minutes.
In Walk-away mode, place card in Alere Reader which will automatically interpret test results at 15 minutes.
Results are displayed on the screen, with a Positive or Negative result. If a test error occurs, “Invalid” displays.
Repeat invalid tests with a new test device.
Call Alere Technical Services if the problem persists.
WALK-AWAY RESULTS WITH ALERE READER
MATERIALS RECOMMENDED
BUT NOT PROVIDED
COLLECT SAMPLE
Visual read interpretation
Alere BinaxNOW® Legionella Antigen Card (12 Test)
852-012
Alere BinaxNOW® Legionella Antigen Card (22 Test)
852-100
Alere BinaxNOW® Legionella Antigen Control Swabs
(5-Pos/Neg Swabs/Kit)
852-010
Click here to learn how to use the Alere™ Reader.
Dip an Alere™ swab into the urine specimen to be tested, completely covering the swab head.
With the Device closed, look through window to interpret results.
Read results immediately. Results read before or after 15 minutes may be inaccurate.
Set time and wait 15 minutes
For a NEGATIVE SAMPLE
• The BLUE Control Line in the BOTTOM THIRD of the window turns a pink to purple color.
No other line appears.
PREPARE REAGENT
www.alere.com
For more information on our products and services, please visit
Hold Reagent A vial vertically, ½ to 1 inch above the card.
Read results visually or with the Alere Reader.
TEMPERATURE
FLU B
© 2019 Abbott. All rights reserved. Alere and BinaxNOW are trademarks of the Abbott group of companies. 120002663-02 09/18
For a FLU A POSITIVE SAMPLE
• The BLUE Control Line in the BOTTOM THIRD of the window turns a pink to purple color.
• A SECOND pink to purple Sample Line appears above it in the MIDDLE THIRD of the window.
Any shade of a pink to purple Sample Line indicates a positive result.
CONTROL
VISUAL INTERPRETATION
READ NOW MODE
A test is INVALID if the Control Line remains BLUE or is not present at all, whether a Sample Line(s) is present or not. Repeat Invalid tests with a new test device.
Call Alere™ Technical Service if the Problem persists.
Allow reagents and cards to equilibrate to room temperature (59–86°F, 15-30°C) before testing. Bring patient urine and/or liquid urine control(s) to room temperature(59–86°F, 15-30°C).
FLU A
After 15 minutes, place developed card in Alere Reader for interpretation. Read time: ~15 seconds.
Results are displayed on the screen, with a Positive or Negative result. Allow Alere Reader to interpret results.
If a test error occurs, “Invalid” displays. Repeat invalid tests with a new test device. Call Alere Technical Services if the problem persists.
For a FLU B POSITIVE SAMPLE
• The BLUE Control Line in the BOTTOM THIRD of the window turns a pink to purple color.
• A SECOND pink to purple Sample Line appears above it in the TOP THIRD of the window.
Any shade of a pink to purple Sample Line indicates a positive result.
ORDERING AND CONTACT INFORMATION
For a FLU A and FLU B POSITIVE SAMPLE
• The BLUE Control Line in the BOTTOM THIRD of the window turns a pink to purple color.
• A SECOND pink to purple Sample Line appears above it in the MIDDLE and TOP thirds of the window.
Any shade of a pink to purple Sample Line indicates positive results.
Do not remove card from pouch until the specimen has reached room temperature
ADD REAGENT
1. Materials
There are two holes on the inner right panel of the card, insert swab into the BOTTOM hole (swab well). Firmly push upwards so that the swab tip is fully visible in the top hole.
2. Temperature
Slowly add three (3) free falling drops
of Reagent A to the BOTTOM hole.
3. Prepare test card
Visual read interpretation – Invalid
ALERE BINAXNOW® LEGIONELLA
URINARY ANTIGEN CARD
4. Collect Sample
With the Device closed, look through window to interpret results. Read results immediately.
Results read before or after 15 minutes may be inaccurate.
BLUE Control Line
REMOVE EXCESS
5. Insert Swab
No Control Line
If the swab drips, touch swab to side of urine container to remove excess liquid.
INTERPRET RESULTS AFTER 15 MINUTES
INSERT SWAB
Read results either visually through the window or with the Alere™ Reader.
Results read beyond 15 minutes may be inaccurate. However, some positive patients may produce a visible specimen line in less than 15 minutes.
Welcome to the
Alere BinaxNOW® Legionella Urinary Antigen Card
Product Demo
EndFragment
This is CLIA Waived for nasal swab and nasopharyngeal swab.
MATERIALS INCLUDED
EndFragment
POSITIVE
The Alere BinaxNOW® Legionella Urinary Antigen Card test kits contain the following:
A positive specimen will give two pink-to-purple colored lines.
This means that antigen was detected. Specimens with low levels of antigen may give a faint Sample Line. Any visible line is positive.
NEGATIVE
A negative specimen will give a single pink-to-purple colored Control Line in the top half of the window, indicating a presumptive negative result. This Control Line means that the detection part of the test was done correctly, but no L. pneumophila serogroup 1 antigen was detected.
If no lines are seen, or if just the Sample Line is seen, the assay is invalid. Invalid tests should be repeated. If the problem persists, contact Alere™ Technical Service.
READ RESULTS
Read results with the Alere™ Reader.
Results may be read with either Read Now or Walk Away mode.
X
6. Use Reagent
TEST DEVICES
A membrane coated with rabbit antibody specific for Legionella pneumophila serogroup 1 antigen and with control antibody is combined with rabbit anti-Legionella pneumophila serogroup 1 antigen and antispecies conjugates in a hinged test card.
REAGENT A
Citrate / Phosphate with Tween® 20 and Azide.
Designed for use with Alere BinaxNOW® Legionella.
Do not use other swabs.
SWABS
L. pneumophilia negative swab.
NEGATIVE CONTROL SWAB
POSITIVE CONTROL SWAB
Heat inactivated L. pneumophilia dried onto swab.
TIMER
Read results with the Alere™ Reader.
Results may be read with either Read Now or Walk Away mode.
INDEX
59–86°F
15–30°C
(+)
or
Invalid result:
(-)